No, matching screens don’t establish that either pulse oximeter is accurate. That comparison gives you an observation about two displays. It doesn’t give you an independent reference for deciding how close either estimate is to the quantity it is meant to estimate.
A paper example, not a home test
Imagine a diagram with two dots marked A and B, one for each device. Put them at the same spot. Now put a third dot marked Reference somewhere else. A and B still agree with each other; both miss the reference in the same direction. Hide the third dot and the first two look exactly the same as before. Their agreement alone cannot tell you where that missing dot belongs.
That is why the comparison in FDA’s March 2013 pulse oximeter guidance, section 4.1.1, is not one fingertip screen against another. For the clinical accuracy comparison described there, FDA recommends pairing the system’s estimate with a simultaneous measurement from arterial blood co-oximetry, and documenting the tested setup and conditions. The two measurements play different roles: one is the estimate being evaluated; the other is the reference used to evaluate it.
Ask what the comparison was made against
If a review says a device “matched our other oximeter,” the fair summary is “the displays agreed in that comparison.” If it calls this proof of accuracy, look for a separate reference study and see what equipment and conditions it actually covers. If there is no reference in the report, leave accuracy unverified by that comparison. You don’t need to repeat it at home to spot the missing evidence.
Sources and review: The dots illustrate the logic of a comparison; they are not patient data. This does not select a device or interpret any reading. Not yet medically reviewed.