Two airflow numbers can have the same units and still answer different questions. Free-air flow describes the compressor without the load of the attached nebulizer. A figure measured with a nebulizer attached describes airflow for the named pairing under the reported conditions. The FDA’s October 1993 reviewer guidance, Performance Requirements – In Vitro, explains why bench reports identify the components tested and describe the setup and parameters.

Put the footnote back in the comparison

Documented flow figures, not a ranking
SourceReported figureWhat belongs next to the number
Drive DeVilbiss U.S. 3655D specification sheet13 L/min, typical free-air flowCompressor free-air figure; this row does not report flow with the bundled disposable nebulizer attached.
PARI Trek S U.S. instructions, Compressor Performance4.0 L/min, operating liter flowThe footnote names an attached PARI LC Sprint or LC Plus reusable nebulizer.

Suppose a comparison sheet labels both rows “system airflow.” Correct the Drive row to “typical free-air flow” and mark paired flow not reported in that sheet. Keep the named attachment with the PARI figure. PARI’s same table also gives an 8.5 L/min maximum; it does not label that row free-air flow or attach the operating-flow footnote to it. Do not fill the missing row with any of those figures.

For a like-for-like documentation comparison, ask each manufacturer for the same measurement with the exact compressor and nebulizer identified, plus the test conditions. If those records aren’t available, leave the paired comparison open. Neither row measures the aerosol that reaches a person or demonstrates that one system is better.