The range a pulse oximeter can display is not necessarily the range covered by its accuracy claim. Check the heading above the number before copying it. In the Nonin Onyx Vantage 9590 manual’s Specifications section, the oxygen saturation display range is 0% to 100%, but the overall rows in the declared non-motion accuracy tables cover 70% to 100%.

Read across one row, then read its heading

In that same manual’s two non-motion tables, the 70% to 100% row reports Arms as ±2 for finger and thumb and ±3 for toe. Copy the site and the non-motion condition with the number. Neither is a body-wide specification. The lower end of the table is part of the reported test range, not a personal target or an action threshold.

FDA’s March 2013 guidance, section 4.1.1, calls Arms a root mean square difference. It summarizes differences from a reference across paired observations. Squaring the differences before averaging keeps errors in opposite directions from canceling. An Arms number is not the largest error seen and the plus-or-minus notation does not promise that every reading lies within that distance of the reference. FDA discusses bias and limits of agreement separately.

Rewrite the claim with its limits intact

Instead of “accurate to two across the whole screen range,” write: “This manual reports ±2 Arms for the finger and thumb row over 70% to 100% in non-motion testing.” For another data sheet, copy five fields before comparing: model and manual revision, statistic as labeled, range in the accuracy row, sensor site and test condition. If the sheet gives only a display range, the accuracy row is still missing.

Sources and review: This reads an identified manual and FDA’s test-document terminology. It is not a product recommendation or a way to judge a personal reading. Not yet medically reviewed.