If a portable nebulizer sat in a warm luggage compartment, look for the row labeled for storage or transport, not the first temperature range you find. The two rows describe different conditions. FDA’s home-use design guidance, sections V.D, V.E and V.J, discusses temperature, humidity and storage separately; it does not supply a single limit for all home devices.

Read the entire row, not just its upper number

For the U.S. OMRON MicroAir NE-U100, section 7 of manual 3608884-7D, printed page 32, gives a storage and transport range of -4°F to +140°F, with 10% to 90% relative humidity, non-condensing. The operating row ends at +104°F and lists its own humidity and air-pressure conditions. Neither number is a universal limit for portable nebulizers.

Worked record: Imagine a logger at the bag location recorded a peak of 110°F, but no humidity. Enter “Peak temperature: 110°F (inside luggage compartment, next to bag); compare with storage/transport row, p. 32; humidity and condensation not documented.” The peak is below the listed storage upper limit and above the operating upper limit. That is a comparison of one recorded value, not evidence that all storage conditions were met or that the unit is fit for use.

If all you have is a city weather reading, note where it came from rather than entering it as the temperature next to the device. Bring the supplier or manufacturer the exact model, the location and duration of the exposure if known, and the measurements you actually have. Leave the disposition unresolved if the record is incomplete. The cited specifications do not tell you how to assess damage or how long to wait after a temperature excursion, and they do not specify medication storage.

Review status: Not yet medically reviewed.