A standards number is not enough to tell you which meter the evidence covers. The FDA recognition entry for ISO 23747 identifies the second edition, August 2015, and describes its scope as peak expiratory flow measurement in stand-alone or integrated equipment. That entry identifies a standard, not a tested product.

Follow the claim to the subject device

  1. Find the edition. In Microlife’s PF200B summary, K253180, section 7, performance-testing item 6, the submitter lists ISO 23747, second edition, August 2015. Record the edition from that entry rather than completing an undated claim from memory.
  2. Find the configuration the summary describes. In sections 2 through 4 of the same summary, PF200B is the subject device, described as a digital meter using a turbine and optical sensing. PF-100 and PF-100-1 appear in the predicate section. A predicate is a comparison device in the submission, not another name for the subject.
  3. Separate the model description from the lab setup. The performance standards list names the edition but is not a specimen-by-specimen account of what was on the bench. If the claim depends on a particular accessory or test setup, note which detail you could not verify in that list and ask for the matching report or configuration reference.

Fictional listing: “PF-100 tested to ISO 23747:2015, see K253180.” Record “edition located; cited subject is PF200B; evidence for the stated PF-100 configuration not established by this reference.” That is a source mismatch, not a finding that the PF-100 failed a test. Nor is the presence of an ISO number itself a clearance decision.

Source and review: A standards trace using the FDA recognition entry and the manufacturer’s public U.S. summary. Not yet medically reviewed.