Several readings can sit close together and still miss the reference used in a lab test. Repeatability asks how close the repeats are to one another under the same conditions. An accuracy assessment asks how close the results are to a reference. In the methods of a 2025 bench study of six peak flow meters, the researchers reported these as separate quantities.
Mark two cells, then find their reference
Open Table 4 and equations 1 and 2 in the original study. Pick one model and one input row, and keep that row fixed as you follow the two columns.
- Repeatability range
- Mark the range. In this paper it is the highest minus the lowest of five readings for that input. It describes the spread; there is no subtraction from the bench reference in that column.
- Measurement error
- Mark the error. In this paper it is the average of the five readings minus the bench reference for that input. It answers a different question about where the readings were centered.
- Scope of the result
- Write “bench input, this row, this tested meter.” The reference was produced on a test bench. The table is not a set of readings from people using meters at home.
Fictional summary: “The repeats cluster tightly, so the reading matches the reference.” Cross out the second half. A narrower description is “the repeats cluster tightly; check the separate error column before making a statement about closeness to the reference.” A report that gives only repeatability leaves that second question unanswered.