Several readings can sit close together and still miss the reference used in a lab test. Repeatability asks how close the repeats are to one another under the same conditions. An accuracy assessment asks how close the results are to a reference. In the methods of a 2025 bench study of six peak flow meters, the researchers reported these as separate quantities.

Mark two cells, then find their reference

Open Table 4 and equations 1 and 2 in the original study. Pick one model and one input row, and keep that row fixed as you follow the two columns.

Repeatability range
Mark the range. In this paper it is the highest minus the lowest of five readings for that input. It describes the spread; there is no subtraction from the bench reference in that column.
Measurement error
Mark the error. In this paper it is the average of the five readings minus the bench reference for that input. It answers a different question about where the readings were centered.
Scope of the result
Write “bench input, this row, this tested meter.” The reference was produced on a test bench. The table is not a set of readings from people using meters at home.

Fictional summary: “The repeats cluster tightly, so the reading matches the reference.” Cross out the second half. A narrower description is “the repeats cluster tightly; check the separate error column before making a statement about closeness to the reference.” A report that gives only repeatability leaves that second question unanswered.