A particle-size number belongs to a particular aerosol test. The FDA reviewer guidance defines mass median aerodynamic diameter (MMAD) as the diameter with half of the particle mass on each side. It does not mean all droplets are that size. Before using a value, match it to the right preparation and test setup.

Find the row rather than copying a headline

The performance tables in K192971 report MMAD in a metric row alongside separate rows for geometric standard deviation and respirable mass. The columns compare the proposed kit with the predicate. Sections 7.3.1 and 7.3.2 use different supplied-air conditions; do not move a value from one table into the other. The preparation also stays with the reported number. Those are test results, not permission to use a particular medicine with a different device. The NE-C801 manual also warns that particle size and nebulization rate can vary with the type of medication, and documents a named sampling method in its test notes.

Write a particle-size citation someone else can reproduce

System
Exact device, cup or delivery configuration when given, document revision, and page: ___.
Test
Named preparation, measurement method, and sampling conditions as written: ___.
Reported result
Exact metric name and row. If the source instead gives a range, fraction or mass, do not rewrite it as MMAD: ___.
Comparison boundary
Which inputs and endpoints would need to match before a comparison could be made? ___.

Worked test entry: For the proposed Nebulizer Kit column in K192971, record the albuterol sulfate MMAD row in section 7.3.1 as belonging to the 8 L/min supplied-air bench condition. The corresponding section 7.3.2 row belongs to the 4 L/min bench condition. Keep them on different lines, even though the metric and named preparation match. These are test conditions, not operating instructions.

For example, if a listing gives an MMAD and a second listing gives a respirable fraction, stop at “different metrics.” A number in micrometers and a percentage do not answer the same question. Even two values with the same unit do not establish that one produces a better outcome.

This article reads published bench data. It does not estimate where medicine deposits for a person, select a setup, or interpret a prescription.

Review status: Not yet medically reviewed.