A single accuracy figure cannot fill a table about performance by skin pigmentation. Keep each finding tied to the model and sensor, the people studied, the way researchers classified skin pigmentation, the comparison method and the range they analyzed. If a report names the model but does not give the subgroup figures you need, leave the cells marked “not shown in this report.”
Three rows from the testing team’s own reports
These Open Oximetry Project reports identify the UCSF Hypoxia Lab as the source of the test data. In the two reports with subgroup values used below, the site describes healthy volunteer cohorts and groups defined using individual typology angle (ITA), an instrument-derived skin-color measure. The table records reported subgroup mean bias: the average oximeter estimate minus the arterial blood co-oximetry reference, in percentage points. It uses the site’s subgroup fields for the 70–84% reference range, not the site’s distinct modeled differential-bias fields. The range is part of the lab analysis, not an action threshold.
| Configuration and direct report | Population and pigmentation field | Mean bias, dark / light |
|---|---|---|
| Masimo Rad-97 with adult reusable sensor | Healthy volunteers; ITA groups. Report lists 88 participants for its overall accuracy cohort. | +4.92 / +2.02 percentage points |
| ChoiceMMed MD300CN350R (2020 report entry) | Healthy volunteers; ITA groups. Report lists 45 participants for its overall accuracy cohort. | +0.48 / +1.31 percentage points |
| ChoiceMMed MD300CN340 | Healthy volunteer lab data; the page describes a forehead skin-color figure using the Monk scale. | Not shown in the inspected report. Its overall accuracy figure is not this metric. |
In these specific entries, the dark-group mean is higher than the light-group mean in the first row and lower in the second. The listed totals are overall cohort sizes, not numbers in either subgroup. These point estimates alone do not show the uncertainty of that contrast, explain its cause or rank the devices. The third row is a gap in this report, not a conclusion that no other evidence exists.
Add another row without borrowing a result
Use this row format: exact device and sensor ___; report link and table or panel ___; participants and setting ___; pigmentation method, site and group definitions ___; reference method and range ___; same outcome for each group ___; subgroup counts or uncertainty if reported ___; what is missing ___. If the report shows only overall error or only a different model, keep it in the notes, not in the subgroup cells.
Sources and review: This table transcribes selected lab results, not Breathwell testing. It is not a prediction for home use or a correction to apply to a reading. Not yet medically reviewed.