
Before comparing two CPAP noise figures, check whether they describe the same sound measure and the same equipment configuration. A listing that supplies only a number followed by dBA leaves out both checks. Ask for the full line from the manual rather than assuming the numbers can be ranked.
One manual gives four figures for the same machine family
The Resmed AirSense 11 Americas guide, Technical specifications: Sound, reports the following values for CPAP mode under ISO 80601-2-70:2015. It reports an uncertainty of 2 dBA for each entry.
| Configuration in the guide | Sound pressure | Sound power |
|---|---|---|
| SlimLine and half-filled HumidAir 11 tub | 27 | 35 |
| SlimLine and side cover | 25 | 33 |
In that table, 25 and 35 are not just numbers for two configurations. They are also different measures. Even a pressure-to-pressure comparison still needs the configuration and stated uncertainty attached. The table does not tell you how loud the machine will sound in your room or how a mask will sound to a bed partner.
Keep a full row for any listing you compare
Record the model, guide revision and page; the measure as named; the value and uncertainty; the test standard and edition; and the components listed for the measurement. If the vendor cannot tell you which row it quoted, mark the comparison as unresolved. No configuration change is suggested by the example.
Sources and scope: Manufacturer declarations only. Breathwell did not measure sound or rank devices. Not yet medically reviewed.